| Tidal
Volume |
10
ml - 300 ml (Newton Valve)
50 ml - 2000 ml (Standard Valve) |
|
| Frequency
|
10
- 85 (cycle/min) |
|
| Frequency
for HFPPV |
60
- 125 (cycle/min) |
|
| I:E
Ratio |
Continuously
variable and dependant on chosen I & E settings |
|
| Gas
Consumption |
Minute
volume plus 0.1 litre/cycle to power fluid logic circuit |
|
| Minute
Volume (litres) |
1.0
- 30.0 |
|
| Inspiratory
Time (secs) |
0.2
- 2.0 (independent and continuously variable) |
|
| Expiratory
Time (secs) |
0.5
- 4.0 (independent and continuously variable) |
|
| Inspiratory
Flow (litres per sec) |
0.25
- 1.0 (independent and continuously variable) |
|
| Inspiratory
Pressure Relief |
60
cmH2O |
|
| Expiratory
Resistance |
2.5
cmH2O/litre |
|
| Respiratory
Manometer |
Range
- 20 to + 100 cmH2O with zero adjust
facility |
|