physical | functional | alarms

Multivent Ventilator

Technical Specification

 
Physical
Size (mm) 525 wide x 247 deep x 305 high (O model)
525 wide x 247 deep x 594 high (E model)
Weight 20.5 kg
Power Source 88-264 VAC, 45-66 Hz, 20-100 psi (140-690 kPa) Oxygen or Air
Gas Consumption 1/10th of Minute Volume Approximately
Battery Life 100 hours minimum from charge
Battery Recharge 1 hour of recharge gives 50 hours use minimum
Circuit Breaker Switch on rear panel
Respiratory Range -20 to 100 cmH20
Manometer with zero adjust facility

Functional
Rate 5-35 bpm
Tidal Volume 100-1200 ml
I:E Ratio 1:1 to 1:5 (1:2 preset on Ether model)
Maximum Respiratory Pressure 10 to 50 cmH20
Inspiratory Pressure Relief 60 cmH20

Audio Visual Alarms
High Airway Pressure 50 cmH2O cannot be muted
Low Airway Pressure & Disconnect 5 cmH2O (PEEP referenced) - Alarm can be muted for 1 minute
Low Battery Allows for 1 hour's use minimum - Alarm can be muted for 4 minutes